LASER THERAPY HAMMER Model: COZING‑T05/T05‑03/04/05/06/07/08/09 Thank you for your trust on our semiconductor laser therapy hammer. Please read the instruction carefully before using. Understand the indications and the use of methods, ensure use safely. The more info. , pls check the manual book contraindication part 【CAUTION】 Critical Precautions 1. Laser Safety Warnings CLASS IIIB/IV LASER PRODUCT ‒ Potential eye and skin hazard. • Never direct the laser beam at eyes, reflective surfaces, or unintended targets (even brief exposure may cause permanent eye damage) • Use protective eyewear rated for the laser’s wavelength if operational exposure is possible. • Discontinue use immediately if skin irritation, dizziness, or abnormal sensations occur (may indicate accidental exposure). 2. Photosensitivity & Medication Warnings Risk of severe reactions when combined with light‑sensitive substances: • Avoid concurrent use with photosensitizing medications (e.g., tetracyclines, retinoids, thiazides, sulfonamides). Consult a physician if uncertain. • Wait ≥48 hours after photodynamic therapy (PDT) or topical photosensitizer use (e.g., 5‑ALA) before laser exposure. 3. Additional High‑Risk Conditions • Do not use on recently tanned skin, active infections, or areas with tattoos/permanent makeup (risk of burns or pigment changes). • Pregnancy, epilepsy, or immunosuppression: Safety not established. Seek medical advice. CONTENTS Caution 01 Product introduction 02 Indications 04 contraindication 06 Specification 08 Operation instruction 10 Maintenance 11 Warranty card 11 【PRODUCT INTRODUCTION】 | Product Name | Laser Therapy Hammer | | --- | --- | | Manufacturer | Wuhan Cozing Medical Devices Co., Ltd. | | Safety Class | Class 3R/Class IV | | Therapeutic Method | Non‑invasive transcutaneous photobiomodulation | | Address | Building 8, Guannan Fuxing Medical Park, No. 58 Optical Valley Avenue, Wuhan | Therapeutic Wavelength Principles : 650nm (Red Light ‑ 10mW per diode) Mechanism: Absorbed by mitochondrial cytochrome c oxidase, enhancing ATP production and cellular repair. Stimulates collagen synthesis (skin rejuvenation) and microcirculation (wound healing). 808nm / 810nm (Near‑Infrared ‑ 200mW to 500mW per diode) Mechanism: Deeper penetration (3‑5cm), targeting muscles, joints, and nerves. Reduces inflammation (modulates COX‑2/PGE2) and pain (endorphin release). Accelerates tissue repair via fibroblast activation. 904nm (Pulsed Infrared ‑ High Peak Power, Low Average Power ≤3.87 mW) Mechanism: Pulsed mode penetrates 5‑7cm, reaching deep joints/bones. Triggers anti‑inflammatory cytokines (IL‑10) and neuroregeneration. Safety: Short sessions (3‑5min) due to high peak power but low thermal risk. Multi‑Wavelength Synergy 650nm + 808nm: Combines superficial healing (red) with deep pain relief (NIR). 904nm Pulsed + Continuous 810nm: Targets acute/chronic pain simultaneously. Key Safety & Usage Notes Pre‑Treatment: Clean skin with non‑alcoholic antiseptic to avoid light scattering. Contact Angle: Maintain 90°±5° for optimal energy delivery. Session Limits: 904nm Pulsed: ≤5min/day (high peak power). High‑Power 808nm (500mW/diode): ≤15min to prevent thermal buildup. Low‑Power 650nm (10mW/diode): Safe for extended use (20‑30min). *(Refer to Table 1 for intensity levels and device‑specific protocols.)* Design Tip:Use icons (e.g., mitochondria for 650nm, bone icon for 904nm) to visually reinforce mechanisms in the manual. COZING‑T05 Series Treatment Parameters Reference Table 【INDICATIONS】 |Model|Time|Frequency|Course|Application| | --- | --- | --- | --- | --- | ||15‑20 min|||·Frozen shoulder ·Muscle strain| |COZING‑T05‑04|10‑15 min|||·Lumbar disc pain| ||20‑30 min|||·Diabetic neuropathy| |COZING‑T05‑05|3‑5 min||2‑3 weeks|| ||||1 time|| |COZING‑T05‑07||||·Ankylosing spondylitis| ||||1 time|·Sacroiliac joint pain| ||25‑40 min|2 times||·Migraine| Treatment Protocol Rationale Acute Pain (COZING‑T05‑03/04) Short course (2‑5 weeks), high frequency (1‑2x daily) Chronic Pain (COZING‑T05/09) Extended course (4‑12 weeks), low frequency (1x daily) Super Pulsed Therapy (COZING‑T05‑05 to 08) Brief sessions (≤10 min) , 48h intervals (avoid cumulative effects) Usage Guidelines Treatment Intervals: Superficial (<3cm depth): ≥6h between sessions Deep tissue (≥3cm): ≥24h between sessions Course Breaks: Pause 1 week after each completed course for reevaluation 【Contraindication】 General Contraindications (All Models) | Absolute Contraindications | Relative Contraindications | Special Notes | | --- | --- | --- | | • Pregnancy (abdominal use) | • Autoimmune diseases (e.g., rheumatoid arthritis, SLE) | • Metal implants/tattoos/dark skin (Fitzpatrick IV‑VI): Reduce fluence, avoid thermal accumulation | | • Active cancer/tumors | • Thyroid dysfunction (critical for NIR models) | • Protocol compliance: Pre‑treatment patch test and continuous patient monitoring required | | • Photosensitive disorders (lupus, porphyria) | • Steroid injections within 72 hours | | | • Epilepsy/seizure disorders | | | | • Active HSV/skin infections | | | | • Photosensitizing medications (e.g., tetracyclines, isotretinoin) | | | | • Open bleeding wounds | | | | • Pacemaker/implantable device areas | | | Model‑Specific Contraindications 1. COZING‑T05‑03/04/09 (Visible/NIR: 650nm + 810nm/808nm) | Absolute Contraindications | Relative Contraindications | Special Notes | | --- | --- | --- | | • Retinal diseases (no ocular exposure) | • Keloid scarring history | • Laser safety goggles mandatory (650nm red light) | | • Chemical peels within 2 weeks | • Rosacea | • Avoid thyroid region: 810nm NIR may interfere with gland function | 2. COZING‑T05‑05/06/07/08 (Super Pulsed: 904nm High‑Power Pulsed) | Absolute Contraindications | Relative Contraindications | Special Notes | | --- | --- | --- | | • Bone growth stimulators<br>• Osteomyelitis (risk of bacterial dissemination)<br>• Unhealed fractures | • Osteoporosis<br>• Joint prosthesis areas | • Professional use only (certified operators)<br>• Continuous probe motion: Prevent static thermal injury (peak power >20W/cm²)<br>• Deep tissue caution: Avoid bone penetration (904nm high tissue penetration) | 3. COZING‑T05‑09 (Blue LED: 450nm) | Absolute Contraindications | Relative Contraindications | Special Notes | | --- | --- | --- | | • Active retinoid therapy (phototoxicity risk)<br>• Retinal disorders (blue light sensitivity) | • Melasma (may exacerbate)<br>• Vitiligo (risk of uneven pigmentation) | • Avoid hyperpigmented areas (e.g., solar lentigines)<br>• Treatment protocol: ≤10 minutes /session, ≥48‑hour intervals | High‑Risk Conditions (Require Medical Clearance) The following conditions necessitate prior physician approval: Bleeding disorders( INR >3.0 or platelets <50 ×10⁹ / L) Diabetic neuropathy/angiopathy(microcirculation impairment risk) Seizure history (last 12 months): Neurologist consultation required Cardiac pacemakers/implantables (contraindicated for high‑power models T05‑05/06/07/08) Active autoimmune diseases (e.g., scleroderma flare‑ups) Power Density & Safety Thresholds: Super Pulsed models (COZING‑T05‑05/06/07/08): Limit dwell time to <30 seconds per point when peak power density exceeds 5W/cm² to prevent dermal burns. Blue LED (T05‑09): Cumulative dose control (<60 J / cm² / session ) due to 450nm oxidative stress induction. Wavelength‑Specific Risks: 810nm/808nm:Penetration depth 4‑6cm; avoid prolonged exposure over endocrine‑sensitive regions (thyroid, pineal gland). 904nm pulsed: Hemoglobin absorption prioritization; contraindicated in coagulopathies. Pediatric Use: Only COZING‑T05‑03/04 (low‑power NIR) permitted under pediatric protocols; all other models prohibited for patients <18 years Regulatory & Safety Compliance Class IIIB/IV Laser Certification: Super Pulsed models (COZING‑T05‑05 to 08) must comply with FDA 21 CFR 1040.10/EU IEC 60825. Class IIIB/IV Laser Certification:High‑power devices require ANSI Z136.3 or equivalent laser safety training. Documentation:Record wavelength, dose, and patient response for all treatments; retain records ≥5 years. 【SPECIFICATION】 1. General Parameters Probe Geometry: Circular (Ø55mm) Treatment Area: 23.76 cm² *(Calculated from (5.5/2)²)* Operation Modes: Continuous/Pulse (1‑9999Hz adjustable) 1.1 Model‑Specific Specifications COZING‑T05 Series Technical Specifications | Model | Wavelength Combination | Total Power | Diode Configuration | Power Density (*5cm² Probe) | | --- | --- | --- | --- | --- | | T05‑3 | 650nm + 810nm | 8W | 15×650nm (5mW) 10×810nm (200mW) | 87.33mW/cm² (650nm: 3.15W/cm² 810nm: 84mW/cm²) | | T05‑4 | 650nm + 810nm | 10.4W | 9×650nm (200mW) 16×810nm (500mW) | 0.44W/cm² (650nm: 0.036W/cm² 810nm: 0.404W/cm²) | | T05 | 650nm + 808nm | 1.29W | 19×650nm (10mW) 6×808nm (200mW) | 0.054W/cm² (650nm: 0.038W/cm² 808nm: 0.016W/cm²) | | T05‑5 | 650nm + 904nm | 550W* (Avg: 3.87mW) | 3×650nm (10mW) 22×904nm SP (25W) | 23.15W/cm²*(Peak) (650nm: 0.006W/cm² 904nm Avg: 0.017W/cm²) | | T05‑6 | 650+808+904nm | 280.5W* (Avg: 5.12W) | 3×650nm (10mW) 11×808nm (500mW) 11×904nm SP (25W) | 11.81W/cm²*(Peak) (Avg: 0.216W/cm²) | | T05‑7 | Multi‑spectral | 130.1W* (Avg: 4.8W) | 10×650nm (10mW) 5×808nm (500mW) 5×904nm SP (25W) 5×980nm (500mW) | 5.38W/cm²*(Peak) (Avg: 0.198W/cm²) | | T05‑8 | 650+904nm | 50.06W* (Avg: 3.2W) | 6×650nm (10mW) 2×904nm SP (25W) | 2.11W/cm²*(Peak) (Avg: 0.135W/cm²) | | T05‑9 | 650+808+450nm | 3.45W | 10×650nm (10mW) 6×808nm (500mW) 7×450nm LED (50mW) | 0.145W/cm² (650nm: 0.02W/cm² 808nm: 0.12W/cm² 450nm: 0.005W/cm²) | Clinical Correlation: Peak power densities >5 W / cm² (T05‑5 to T05‑7) require movement during application to avoid localized heating SP = Super Pulsed Laser (μs‑level pulses with kW peak power)* 904nm average power is 3.87mW 5. Safety Features Automatic Shutdown: Temperature >42°C (probe surface) Pulse overload (>110% rated power) Beam Safety: 5‑second delay on startup Emergency stop button (mechanical cutoff) 7. Model Comparison Guide Wavelength Penetration Depth: 450nm: 0.2‑0.5mm (Superficial) 650nm: 2‑8mm (Dermal) 808nm: 3‑5cm (Muscle) 904nm SP: 5‑10cm (Deep Tissue) Typical Applications: T05‑3/T05‑4: Musculoskeletal rehabilitation T05‑5/T05‑6: Chronic pain management T05‑9: Dermatological conditions ADDED BATTERY SPECS: Lithium‑ion battery: 18650 cell, 5200mAh capacity(only COZING‑T05 has the battery inside, the other models no battery inside) 【OPERATION INSTRUCTION】 Power Management: Long press (3s): System ON/OFF Short press: Start/pause therapy Parameter Adjustment Matrix: Operation Guide: Power On/Off: Hold the power button for 3 seconds. A beep will sound and the screen will light up. Navigation: Press "OK" to move a yellow cursor through the menu options. Frequency (F): With the cursor on F(Hz), short press +/- to change by 1 Hz; hold to change by 10 Hz。 The max. F is 9999Hz. Recommended: 10Hz/40Hz. Time: With the cursor on TIME, short press +/- to change by 5 seconds; hold to change by 5 minutes. Wavelength (WL): With the cursor on WL, press +/- to cycle through the available wavelengths. Power Intensity: With the cursor on POWER, press +/- to adjust through 10 intensity levels. Laser Output: Short press the power button to start output. Adjusting settings pauses the laser. It resumes automatically after 6 seconds, or immediately with a short press of the power button. SAFETY INTERLOCKS ADDED: Automatic shutdown at >40°C casing temperature Emergency stop: hold the power button for 3 seconds 【MAINTENANCE】 SCHEDULED SERVICE REQUIREMENTS: Daily:Output window cleaning with soft cloths COZING‑T05 Battery Maintenance & Safety(The other models without battery) 1.Always use the original charger to avoid overcharging. 2.For long‑term storage, keep the battery charge between 40%‑60% and store in a cool, dry place. 3.Do not disassemble the battery or expose it to extreme temperatures. Stop using immediately if swelling or leakage occurs. REVISED PACKING LIST: | Main Unit | | --- | | Charger | | Manual | 【Warranty Card】 Thank you very much for using COZING‑T05. Customer name Factory Serial number : ROHS CE PRODUCT: INSOMNIA TREATMENT APPARATUS Model:COZING‑T05 series Tel:0086‑27‑592 617 71 Address: Building 8 Guannan Fuxing Medical Park, No.58 OpticalValley Avenue,East Lake High‑tech Development Zone,Wuhan,Hubei,China
【CAUTION】
Critical Precautions
1. Laser Safety Warnings
CLASS IIIB/IV LASER PRODUCT ‒ Potential eye and skin hazard.
• Never direct the laser beam at eyes, reflective surfaces, or unintended targets (even brief exposure may cause permanent eye damage)
• Use protective eyewear rated for the laser’s wavelength if operational exposure is possible.
• Discontinue use immediately if skin irritation, dizziness, or abnormal sensations occur (may indicate accidental exposure).
2. Photosensitivity & Medication Warnings
Risk of severe reactions when combined with light‑sensitive substances:
• Avoid concurrent use with photosensitizing medications (e.g., tetracyclines, retinoids, thiazides, sulfonamides). Consult a physician if uncertain.
• Wait ≥48 hours after photodynamic therapy (PDT) or topical photosensitizer use (e.g., 5‑ALA) before laser exposure.
3. Additional High‑Risk Conditions
• Do not use on recently tanned skin, active infections, or areas with tattoos/permanent makeup (risk of burns or pigment changes).
• Pregnancy, epilepsy, or immunosuppression: Safety not established. Seek medical advice.
【PRODUCT INTRODUCTION】
| Product Name | Laser Therapy Hammer |
| --- | --- |
| Manufacturer | Wuhan Cozing Medical Devices Co., Ltd. |
| Safety Class | Class 3R/Class IV |
| Therapeutic Method | Non‑invasive transcutaneous photobiomodulation |
| Address | Building 8, Guannan Fuxing Medical Park, No. 58 Optical Valley Avenue, Wuhan |
Therapeutic Wavelength Principles :
650nm (Red Light ‑ 10mW per diode) Mechanism:
Absorbed by mitochondrial cytochrome c oxidase, enhancing ATP production and cellular repair.
Stimulates collagen synthesis (skin rejuvenation) and microcirculation (wound healing).
808nm / 810nm (Near‑Infrared ‑ 200mW to 500mW per diode) Mechanism:
Deeper penetration (3‑5cm), targeting muscles, joints, and nerves.
Reduces inflammation (modulates COX‑2/PGE2) and pain (endorphin release).
Accelerates tissue repair via fibroblast activation.
904nm (Pulsed Infrared ‑ High Peak Power, Low Average Power ≤3.87 mW) Mechanism:
Pulsed mode penetrates 5‑7cm, reaching deep joints/bones.
Triggers anti‑inflammatory cytokines (IL‑10) and neuroregeneration.
Safety: Short sessions (3‑5min) due to high peak power but low thermal risk.
Multi‑Wavelength Synergy
650nm + 808nm: Combines superficial healing (red) with deep pain relief (NIR).
904nm Pulsed + Continuous 810nm: Targets acute/chronic pain simultaneously.
Key Safety & Usage Notes
Pre‑Treatment: Clean skin with non‑alcoholic antiseptic to avoid light scattering.
Contact Angle: Maintain 90°±5° for optimal energy delivery.
Session Limits:
904nm Pulsed: ≤5min/day (high peak power).
High‑Power 808nm (500mW/diode): ≤15min to prevent thermal buildup.
Low‑Power 650nm (10mW/diode): Safe for extended use (20‑30min).
*(Refer to Table 1 for intensity levels and device‑specific protocols.)*
Design Tip:Use icons (e.g., mitochondria for 650nm, bone icon for 904nm) to visually reinforce mechanisms in the manual.